START-UP 

a)  Remove ALL PLASTIC WRAPPING AROUND FILTERS PRIOR TO USE. 

b) 

Ensure that supply and return air grilles are not obstructed in any way (air circulation patterns will be inhibited if airflow is obstructed); 

c) 

Insert male end of cord into 15-amp circuit.  Unit should be connected to an independent circuit; 

d) 

Turn power toggle switch to "ON" position; 

e) 

Plug in remote 

OPTIONAL 

ON/OFF switch and switch toggle to Remote, for remote operation. 

WARNING: DO NOT OPERATE UNIT UNLESS ALL FILTERS ARE IN PLACE. 
 

NOTE: In certain cases 

4.0 

 

PRINCIPLE OF OPERATION 

The 

DUCTLESS FUME HOOD, DFH-1000

 features twofold method of operation; 

 

a)  captures general particulate; 

b) 

chemically adsorbs, reacts or scrubs nuisance gases or odors; 

5.0 

 

SYSTEM COMPONENTS 

FILTERS: a) Dust Filters; b) Specially selected chemical filters; c) Dust Filters; 
BLOWER ASSEMBLY: 

The blower is factory tested to ensure quiet, vibration-free operation. 

CONTROLS: Features an on/off switch, to control motor operation, a speed control and a remote for 

OPTIONAL

 remote control. 

PLUG:   Six-foot grounded extension cord 

 

5.1 

UNIT SPECIFICATIONS 

 

DFH-1000-B* 

DFH-1000-M* 

DFH-1000-R* 

DFH-1000-D* 

Airflow 

350 cfm 

350 cfm 

350 cfm 

450 cfm 

Stage 1 Filters: 

(1) F007 

(1) F007 

(1) F007 

(1) F007 

Stage 2 Filters: 

(1) F033-GPC 

(1) F033-GPC 

(1) F033-GPC 

(1) F022 

Stage 3 Filters: 

(1) F033-APC 

(1) F033-GPC 

(1) F033-BPC 

 

Stage 4 Filters 

(1) F022 

 

(1) F022 

 

Amperage:  

1.6 (200watts) 

1.6 (200watts) 

1.6 (200watts) 

1.6 (200watts) 

 

* All of the above filter sequences are installed in the airflow direction (bottom to top).  

Dimensions: Ht: 19.5" (plus 6” hood), Length: 23" Width: 13.5"      Voltage: 120/1/60      Approximate Weight: 50 lbs. 

6.0 

 

EQUIPMENT MAINTENANCE PROCEDURES 

Proper  maintenance  is  critical  to  extend  the  life  of  the  filtration  system.    The  information  presented  below  outlines  basic  maintenance  procedures 
ensuring  the 

DUCTLESS  FUME HOOD, DFH-1000

 unit will provide trouble-free operation for years to come.  The 

DUCTLESS FUME HOOD, DFH-

1000

 is designed to allow quick access to the particulate filters, chemical filters, blower/motor and control panel assembly. 

 

6.1 

GENERAL FILTER MAINTENANCE 

It  is  very  difficult  to  predetermine  a  specific  maintenance  schedule  as  the  rate  of  dust  loading  and  chemical  filter  consumption  will  vary  or  each 
application.  Periodic  inspection  of  the  filters  and  pressure  gauge  during  the  first  few  months  of  operation  should  help  establish  an  appropriate 
replacement schedule. 
QUATRO Air Technologies through its years of experience has established the following recommendations for a "typical" replacement filter schedule: 
 

Filter Type: 

 

 

Part No.:  Qty: 

Suggested Replacement: 

 

Dust Filters 

 

 

F007 

 

every two-three months 

 

Chemical Filters 

 

F033-GPC 

every twelve months (depending on application) 

 

Chemical Filters 

 

F033-APC/BPC 

every twelve months (depending on application) 

 

Hepa Filter 

 

 

F022 

 

Used on DFH-1000-B & D & R

, replace every 24 months

 

Note: 

These recommendations are for average room conditions. 

 

6.2 

PARTICULATE FILTER REPLACEMENT 

a) 

Open with appropriate tool (Phillips screw driver) to avoid stripping of screws; 

b) 

Carefully slide out filters along their support channels, slide clean filters gently into place; 

c) 

If encountering difficulties, confirm that there are no obstructions in the filter track. 

 

6.3 

CHEMICAL FILTER TRAY (F033) MAINTENANCE 

It  is  essential  that  the  chemical  filter  be  replaced  approximately  once  every  twelve  months 

OR

  immediately  following  any  detection  of  odor  on  the 

discharge of the system. 
After  an  operating  period  of  twelve  months  (or  when  the  filters  are  consumed),  call  your  distributor  to  purchase  a  new  set  of  chemical  filtration 
assemblies (Part No: F033-GPC & F033-APC) prior to removal of existing one. 

 