2.4.4 Temperature Calibration
The temperature should be recalibrated every time if the handle,
heating element or soldering tip is replaced.
1) Set the temperature 350
℃
.
2) When the temperature is stable, measure the temperature of the tip
with the thermometer, write down the value.
3) Take out the stopple in the CAL hole, use a “-” or “+” type
screwdriver to adjust the screw in the CAL hole until the thermometer
shows the station’s set temperature (350
℃
)
.
4) Temperature will increase if the screw is turned clockwise and will
decrease if it is turned counter-clockwise.(put the stopple back to the
CAL hole
)
Note: It is recommended to use QUICK 191/192 series temperature
tester to measure tip temperature.