12
Long press the " " and " " button to save.
Note
: It is recommended to use QUICK 191/192 series temperature tester to measure tip
temperature.
11.2 Hot Air Temperature Calibration
1) Set the hot air temperature to 300℃ calibration.
2) Press and hold “ ” and
“
-
”
at the same time to enter the hot air temperature calibration
mode.
3) When the temperature is stable, use the hot air temperature tester to measure the hot air
temperature, press “+” or “–”.
Change the hot air temperature, press and hold "+" and "-" to confirm, and return to the
main screen.
Note: It is recommended to use QUICK 196 series hot air temperature tester to measure hot
air temperature. If there is no QUICK196, it is recommended that the temperature tester be
connected to the temperature head of the sensor and placed 3 to 5mm away from the air
nozzle to measure the temperature.