Sofia 2 may be set to two different modes (WALK AWAY and READ NOW). The procedures for each mode are described below. 

Note:

 The Supervisor has 

the option to set the Sofia 2 to Locked WALK AWAY mode, which will prevent an operator from selecting any other mode than WALK AWAY for running a 
test. 

     

 

In WALK AWAY Mode, the user 

immediately

 inserts the Test Cassette into Sofia 2. 

Sofia 2 will automatically time the test development, and the results will be 
displayed in 15 minutes. 

 

 

 

 

 

 

 

The user places the Test Cassette onto the counter or bench top for 15 minutes (outside of 
Sofia 2) and manually times this development step. Then, the user inserts the Test Cassette 
into Sofia 2. In READ NOW Mode, Sofia 2 will scan and display the test result within 1 minute. 

Warning: Results must not be interpreted past 30 minutes after inoculation. Using the Sofia 
2 past this time may result in false results. 

 

 

 

 

Allow the test to develop for the FULL 15 minutes BEFORE placing it into Sofia 2. 

 

WALK AWAY Mode 

READ NOW Mode 

Using Sofia 2

 

Summary of Contents for Lyme FIA Sofia 2

Page 1: ...irst Graduation of Specimen using the Graduated Pipette provided in the kit Note For Specimens in Transport Media preserved collect 100 μL 2nd Graduation using the Graduated Pipette provided in the kit Transfer the Specimen to the Specimen Diluent Tube and mix the solution by squeezing and releasing the top bulb of the Graduated Pipette 3 times Remove the Graduated Pipette from the Specimen Diluen...

Page 2: ...fia 2 will automatically time the test development and the results will be displayed in 15 minutes The user places the Test Cassette onto the counter or bench top for 15 minutes outside of Sofia 2 and manually times this development step Then the user inserts the Test Cassette into Sofia 2 In READ NOW Mode Sofia 2 will scan and display the test result within 1 minute Warning Results must not be in...

Page 3: ...d Walk Away mode users will only be able to use the walk away mode option for running tests Read Now Mode will be grayed out when in Locked Walk Away Mode 5 Sofia 2 will start and display the progress In WALK AWAY Mode the test results will be displayed after 15 minutes In READ NOW Mode the test results will be displayed on the screen within 1 minute See Interpretation of Results section When the ...

Page 4: ...or the rapid qualitative detection of a Campylobacter specific antigen in human fecal specimens Sofia 2 Campylobacter FIA is designed to detect C jejuni C coli C lari and C upsaliensis from patients with signs and symptoms of gastroenteritis The test is intended for use with preserved fecal specimens in transport media and unpreserved fecal specimens Test results should be considered in conjunctio...

Reviews: