6
OPERATING INSTRUCTIONS
Preparations for the use of the QuikRead QuikLink
Use the QuikRead QuikLink Setup Wizard to configure the desired QuikLink
settings for your QuikRead Instrument. See the QuikRead QuikLink Setup
Wizard instructions for use.
Note that after servicing the QuikRead Instrument or the bar code reader, the
devices need to be reconfigured to the customised settings.
Measurement on QuikLink mode with bar code reader
These instructions for use describe the mode in which the assay results are
sent to a remote LIS/HIS computer. The same procedure is also valid in the
mode in which the results are sent to a printer.
START-UP
•
Switch on the QuikRead Instrument.
•
Check that the LCD displays:
QuikLink mode
Operator ID enabled
or
QuikLink mode
Operator ID disabled
•
Use a magnetic card to set the analyte. The LCD will alternately display:
Ready for use
<test> lot xxxx
and Ready for use
Read patient ID
Note that the magnetic card can only be read when the LCD displays
Read the card
or
Ready for use
.
USE OF THE BAR CODE READER
See the supplied bar code reader user manual for detailed instructions for use.
Use the supplied interface cable to connect the bar code reader to the
QuikRead Instrument. The cable is connected to the bar code reader con-
nector located at the back of the QuikRead Instrument. See Figure 2 for the
connections.
To read a bar code, aim the reader at a bar code and pull the trigger in the
handle. Code scanning is performed along the center of the light bar emitted
from the reading window. This bar must cover the entire code horizontally. Do
not aim the light bar direct at your eyes.
The manufacturer’s warranty for the bar code reader covers defects in ma-
terials or workmanship for 5 years from the date of shipment, provided that
the product is operated under normal and proper conditions. In the event of
a reader malfunction, contact your local supplier to receive information of the
authorised service.