4-20 Repair/Replacement and Calibration
MedTrack CR60 Rehabilitation Treadmill Service Manual
7. Restore power and turn on circuit.
8. Operate the treadmill to verify that it reaches the upper and lower
grade limits (15% and 0%).
Field Functional Test
Ensure that the grade will traverse from minimum to maximum and back.
Walking Deck Assembly
Replacing Rollers
1. Remove the power cords and treadmill hood as described in
“Removing the Treadmill Hood” on page 4-3. Remove optional
siderails.
2. Remove the poly-V drive belt as previously described in “Replacing
the Poly-V Drive Belt” on page 4-8.
3. Remove the end caps from both sides of the belt at the rear of the
walking platform.
4. Remove the siderail trim strips.
5. Remove the four ½-inch hex bolts (two on each side) that hold the
two front roller retainers to the siderail.
6. Slide the front roller up and out from the walk belt.
7. Pull the walk belt towards the rear of the deck, then slide the rear
roller out from between the siderails toward the rear of the treadmill.
Need drawing of CR
Front Roller
Poly V Drive Belt
Siderail Trim Strips
Walk Belt
Rear Roller
End Caps