MedTrack CR60 Rehabilitation Treadmill Service Manual
Repair/Replacement and Calibration 4-25
Removing the Enclosure
1. Turn the treadmill power off and disconnect the power cord from the
power source.
2. Remove optional siderails if present.
Caution!
While performing steps 3-7, hold the enclosure securely,
so that it does not fall while you are removing the screws
and the handrail. Be careful not to scratch the uprights.
3. Use a 1/8-inch Allen wrench to remove the six socket screws from the
rear cover of the enclosure.
4. Use a 3/16-inch Allen wrench to remove the two socket screws near
the handrail.
5. Slide the cover down the uprights to expose the PCBA.
6. Disconnect the controller cable, located on the left side of the PCBA.
7. Remove the hex nut that holds the ground wires to the chassis, then
remove the wire.
8. Cut the cable tie that holds the cable to the controller.
9. Use a 5/32-inch Allen wrench to remove the four socket-head screws
that attach the controller assembly to each upright.
10. Lift the entire controller assembly clear of the uprights.
11. Reassemble following steps 2-10 in reverse order.
Field Functional Test
To verify that the treadmill is operating properly, perform Field Test No.
5. See page D-6 for specific instructions.
Replacing the PCBA
1. Turn the treadmill off and disconnect the power cord from the outlet.
2. Use a 1/8-inch Allen wrench to remove the six screws from the rear
panel of the enclosure.
3. Use a 3/16-inch Allen wrench to remove the two socket screws near
the handrail.
Socket screw
on enclosure
10-24 x 0.50L
Flat Washer