MedTrack CR60 Rehabilitation Treadmill Service Manual
Repair/Replacement and Calibration 4-11
4. Discard the old transformer. Replace with a new one repeating steps 2
and 3 in reverse order.
5. If no other service is required, replace the treadmill hood as described
in step 10, page 4-5.
Field Functional Test
To verify that the treadmill is operating properly, perform “Field Test No.
2” on page D-5.
Replacing the Grade Motor
♦
Place a clean sheet of cardboard or a clean rag on the treadmill deck
before starting this procedure.
♦
Do not elevate the treadmill.
1. Block the treadmill headframe securely with wooden blocks to ensure
that the treadmill will not drop when you remove the grade motor.
2. Remove the power cords and treadmill hood as described in
“Removing the Treadmill Hood” on page 4-3.
Transformer