4-26 Repair/Replacement and Calibration
MedTrack CR60 Rehabilitation Treadmill Service Manual
4. Slide the rear cover down the uprights.
5. Remove the hex nut that attaches the PCBA groundwire, then remove
the wire.
6. Unplug the control cable from the PCBA.
7. Remove the Phillips screws that hold the PCBA to the key panel.
8. Lower the PCBA, then unplug the ribbon cable connecting it to the
key panel.
9. Remove the PCBA from the enclosure.
10. Replace the PCBA and reassemble following steps 3-10 in reverse
order.
Configuring the Controller
When you replace a controller, you must configure it to work with the
specific treadmill.
Perform the following steps to configure the treadmill.
1. Enter the service mode by simultaneously pressing
Stop Belt
,
Slower
,
and
Faster
.
2. Choose the appropriate configuration number from the following:
3. Hold down
Stop Belt
and press the
+
or
-
key until the correct
configuration number appears in the center display.
4. Press
Select
to store the configuration.
5. Exit the service mode by simultaneously pressing Stop Belt, Slower,
and Faster.
Field Functional Test
To verify that the treadmill is operating properly, perform Field Test No.
1. See page D-4 for specific instructions.
Removing the Key Panel
To remove the key panel:
1. Remove the PCBA following the above procedure.
2. Remove the 10 hex nuts that hold the key panel to the enclosure.
3. Lift the key panel off the enclosure.
Treadmill Configuration
Configuration No.
1-12 MPH
CP3
0.6-7.2 MPH
CP1
No configuration
CP- -