MedTrack CR60 Rehabilitation Treadmill Service Manual
Field Functional Tests D-1
D
Field Functional Tests
WARNING! The drive board on the AC Drive Module generates high
voltage that is present whenever the red LEDs are
illuminated: Never touch the drive board when the red
LEDs are on. Bleeder resistors on the AC Drive Module
bleed off the high voltage in approximately two minutes.
Do not touch the drive board until the red LEDs have gone
out.
These procedures apply to the MedTrack CR60. You can perform the
complete test or only the tests required after repair or replacement of parts
as specified in Chapter 4, Repair/ Replacement and Calibration. Please
read the entire chapter before starting. The appropriate test must be
performed as verification whenever any of the procedures in Chapter 4
are used.
Complete Field Functional Test
The following steps must be performed with the treadmill completely
assembled, except as noted.
With the power cord connected to the correct outlet, turn on the circuit
breaker.
Test A: Test the Controller Displays
Observe the controller displays during initialization. For the MedTrack
CR60, the normal sequence is:
•
Seven segment displays show all 8s (888 8888 888).
•
After 3 seconds, all displays go blank.
•
Multi display briefly shows configuration code (0.0 CP.-x 0.0).
•
Normal display is shown (00 150 00).
If the normal sequence does not occur, refer to the following table for
possible causes.