2014-09-30
Page 31 of 34
R-Biopharm AG
Technical specifications
RIDA
®
CUBE SCAN Instrument
Photometer resolution
0.0001 ABS
Reproducibility
<1.5% CV at 1 OD
Linearity
0.1000 – 3.0000 OD better than +/- 1.5% and +/- 0.01 OD
Temperature control
Electrical temperature control of the photometer unit to 37°C +- 2°C
Fuse
2.5 A, self-healing
Dimensions
160 x 135 x 145 mm (H x W x D)
Weight
2.2 kg (unpackaged)
Communications interface
RS232, USB, Bluetooth
Tolerance conditions:
Work space:
20 - 28°C; relative humidity: 10 – 85%
Transport/Storage: 0 - 50°C; relative humidity: 5 – 85%
Work surface:
A dry, clean, level surface. Avoid direct sunlight.
Power usage:
12V DC, 2A
Power supply
Manufacturer
Globtek
Type
GTM21097-5012
Mains adapter
A separate AC to DC mains adapter with double insulation
Input
90-264V AC, 47-63 Hz
Output
12V DC, 4.17A
Power usage
Max. 30 VA
Options
Thermo printer
Seiko DPU-414
Interface
Serial
Mains adapter
100-240 VAC
Barcode reader (scanner)
Datalogic Touch65
Reading area
63 mm
Max. resolution
0.10 mm (4 mils)
Mains adapter
100-240 VAC