Fitting Instruction – stabilisers – anti tip (Access / Accent 01/01/2013
onwards)
Part number
MSAWU1127
Where used
Access & Accent Self Propelled, Attendant Propelled
Curved stabiliser can be fitted with the wheel downwards and close to the
floor, or upwards, method of fitting to be assessed dependent upon users and
intended environment.
Tools and Techniques
The following tool will be required;
10mm A/F Open End Spanner
Damaged or worn, components need to be replaced, if assembled in pairs
Remploy Healthcare recommends replace both assemblies, for continuity of
wheelchair in use
.
1. Refer to stabiliser configuration table and check to see if fitted
extension tube (8) is required.
2. If required move to point 3, if not required remove wheels (4) by
lo0sing bolt (6) and nut (3). Remove extension tube (8) by losing bolt
(7) and nut. Refit wheels (4) onto tube (2) by using bolt (6), washer (5)
and nut (3) through hole 10. Tighten nut and bolt.
3. Fit Stabilisers (with or without extension tube (8)) inside bottom rail of
wheelchair and secure using Pin (1) through hole in wheelchair bottom
rail and hole (11) in anti-tipping stabiliser.
4. Ensure stabilisers stop wheelchair from tipping over backwards in fitted
position before use.