1260-100/101 Adapt-a-Switch Carrier Module User Manual
Racal Instruments
1998
Maintenance 4-1
Chapter 4
MAINTENANCE
PERFORMANCE VERIFICATION PROCEDURE
Purpose
The following procedure verifies the basic functions of the
1260-100/101 Adapt-a-Switch Carrier. Test coverage includes:
•
Option-01T power-on self-test
•
Option-01T basic operation in message based mode
•
Option-01T operation in register-based mode
•
1260-100/101 receive and transmit functions
Equipment
Required
•
VXIbus chassis
•
Embedded slot 0 controller, or MXI slot 0 controller with
IBM-compatible PC
Software Required
•
Microsoft Windows 95 or NT
•
VXI Resource Manager (RESMAN) by National Instruments
•
VXI Interactive Control (VIC) by National Instruments
Procedure
1. Ensure that an Option-01T is installed in the carrier as
outlined in the
Option-01T
section of
Chapter 2: Getting
Started
.
2. Set the logical address switch to 3.
3. Set the module address switch to the 1-6 position.