1261B User Manual
Maintenance 7-10
Racal Instruments
Ó
1995
Figure 7-6, Removing The Power Supply
4.
Firmly grasp each handle and pull evenly using a
slight side to side rocking motion. If the power
supply is difficult to remove, recheck to make sure
all the thumbscrews are disengaged.
5.
To reinstall a power supply into the mainframe, first align
the power supply with the cavity at the rear of the
mainframe. Note that the silkscreen on the supply rear
panel will read correctly when the supply is in the proper
orientation. Use the alignment of the threads on the upper
fasteners as a visual guide.
6.
Gently guide the power supply forward until
engagement of the blind-mate connector is felt.
7.
Exert pressure by using the palm of hand to assist
final engagement.
8.
When fully seated tighten the six thumbscrews.
Do not over-tighten.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com