1261B User Manual
Using The Enhanced Monitoring System 4-14
Racal Instruments
Ó
1995
STAT:QUES:TIME:COND?
-- Elapsed Time Condition
STAT:QUES:VXI:COND?
-- VXI BERR/IACK Condition
The bit assignments for the various Condition Registers are
shown in
Tables 4-1
through
Tables 4-6.
Table 4-2, Voltage Condition and Event Register Bit Assignments
Bit
Set When
0
+5V Power Supply Voltage out of tolerance
1
-5.2V Power Supply Voltage out of tolerance
2
-2V Power Supply Voltage out of tolerance
3
+24V Power Supply Voltage out of tolerance
4
-24V Power Supply Voltage out of tolerance
5
+12V Power Supply Voltage out of tolerance
6
-12V Power Supply Voltage out of tolerance
Table 4-3, Current Condition and Event Register Bit Assignments
Bit
Set When
0
+5V Power Supply Current out of tolerance
1
-5.2V Power Supply Current out of tolerance
2
-2V Power Supply Current out of tolerance
3
+24V Power Supply Current out of tolerance
4
-24V Power Supply Current out of tolerance
5
+12V Power Supply Current out of tolerance
6
-12V Power Supply Current out of tolerance
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com