1261B User Manual
Racal Instruments
Ó
1995
Using The Enhanced Monitoring System 4-19
The Enable Registers for each attribute may be set with the
following commands (arbitrary register values are shown)
STAT:QUES:VOLT:ENAB 127
-- Voltage Enable
STAT:QUES:CURR:ENAB 31
-- Current Enable
STAT:QUES:TEMP:ENAB 7
-- Temperature Enable
STAT:QUES:TIME:ENAB 15
-- Elapsed Time Enable
STAT:QUES:FREQ:ENAB 15
-- Fan Speed Enable
STAT:QUES:VXI:ENAB 255
-- VXI Enable
Each of the monitored attributes have an Event Register and an
Enable Register. For each attribute, the registers are logically
ANDed together, and the result is Ored to form a status
summary bit. Each attribute is represented in the “Questionable
Condition” register by a single summary bit. The bit assignments
for the “Questionable Condition” register is shown in
Table 4-8
.
Table 4-8, Questionable Condition Register Bit Assignment
Bit
Set When
0
Voltage Event AND Enable bit or bits set
1
Current Event AND Enable bit or bits set
2
Time Event AND Enable bit or bits set
3
Not Assigned/always 0
4
Temperature Event AND Enable bit or bits set
5
Fan Speed Event AND Enable bit or bits set
6
Not Assigned/always 0
7
Not Assigned, always 0
8
Not Assigned, always 0
9
VXI Event AND Enable bit or bits set
10
Not Assigned, always 0
11
Not Assigned, always 0
12
Not Assigned, always 0
13
Not Assigned, always 0
14
Not Assigned, always 0
15
Not Assigned, always 0
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com