Using The Enhanced Monitoring System 4-27
To determine which line is presently connected, the query:
VXI:CONF:MONITOR:TRIG:OUTPUT?
may be used. The reply to this query will be one of the following:
TTLT0
TTLT1
TTLT2
TTLT3
TTLT4
TTLT5
TTLT6
TTLT7
NONE
INP
DFI
The Default state for this command is “NONE”.
Discrete Fault
Indicator
Discrete Fault Indicator (DFI) allows the user to either connect
an external alarm, eg. A Sona-Alert, close an external relay to
sound a larger alarm or turn on an emergency flashing light if the
1261B chassis EMS module senses an alarm condition.
When enabled the DFI would output a true on the “TRIG
OUTPUT TTL” line (low), J413-2, when any of the following
errors occur:
1) Over Current on any rail.
2) Over Voltage on any rail.
3) Under Voltage on any rail.
4) Over Ambient Temperature (55
°
C).
5) Over Slot Temperature maximum (55
°
C) of any slot.
6) Fans below minimum fan speed of 2000 RPM.
Programming the
Discrete Fault
Indicator
The default mode is DFI OFF after power up or RESET. To
enable the DFI operation send:
VXI:CONF:MONITOR:TRIG:OUTPUT DFI
To confirm the DFI is on send:
VXI:CONF:MON:TRIG:OUTPUT?
Reply is “DFI”