1269 User Manual
SCPI Command Reference 5-4
MEASure:VOLTage1?
-- read +5V Power Supply
MEASure:VOLTage2?
--read -5.2V Power Supply
MEASure:VOLTage3?
-- read -2V Power Supply
MEASure:VOLTage4?
--read +24V Power Supply
MEASure:VOLTage5?
-- read -24V Power Supply
MEASure:VOLTage6?
--read +12V Power Supply
MEASure:VOLTage7?
-- read -12V Power Supply
Each of these has a long form and a short form. Thus
MEAS:VOLTAGE2?
MEAS:VOLT2?
are equivalent.
Omitting the keyword suffix selects the same capability as
specifying the suffix of 1. This means that:
MEAS:VOLTage?
MEAS:VOLTage1?
are equivalent.
The MIN, MAX, and
DEF Parameters
Substitute MINimum, MAXimum, or DEFault in place of a
numeric parameter for most commands. For example, the
following command:
:VOLTAGE3:RANGE:UPPER MIN
sets the upper voltage range for the -2V power supply monitor
to its minimum programmable value. The command:
:VOLTAGE3:RANGE:UPPER DEF
sets the upper voltage range for the -2V power supply monitor
to its default, power-on value.
The MINimum, MAXimum, and DEFault keywords are accepted
for all SENSE:<Attribute>:RANGE commands and the
VXI:CONF:MON:TRIG:DELAY command.
Querying
Parameter Setting
Query the current value of most parameters by adding a
question mark ( ? ) to the command. For example, the following
command sets the trigger delay to 0.5 seconds: