1269 User Manual
Specifications A-2
Table A-2, DC Output Power
Characteristic
Description
Maximum Power Available
To Modules
720 W, 0
o
C to 45
o
C,
Derate 10W/
o
C above 45
o
C
DC Current Capacity (I
MP
)
Voltage I
MP
(Steady-State Current)
+24V 4 A
+12 V 8 A
+5 V 54 A
-2V 12 A
-5.2 V 12 A
-12 V 8 A
-24 V 4 A
+5 V STDBY 0 A (1 A max, user supplied)
DC Voltage Regulation
Voltage Tolerance,
ªª
V
+24 V +1.2V, -720 mV
+12 V +600 mV, -360mV
+5V +250 mV, -125 mV
-2V -100 mV, +100 mV
-5.2 V -260 mV, +156 mV
-12 V -600 mV, +360 mV
-24 V -1.2 V, +720 mV
Maximum Load
Ripple/Noise
(per VXIbus 1.4)
Voltage Ripple/Noise
+24 V 150 mV
p p
+12 V 50 mV
pp
+5 V 50 mV
pp
-2 V 50 mV
pp
-5.2 V 50 mV
pp
-12 V 50 mV
pp
-24 V 150 mV
pp
Maximum Induced
Ripple/Noise
Voltage Ripple/Noise
+24 V 150 mV
p p
+12 V 50 mV
pp
+5 V 50 mV
pp
-2 V 50 mV
pp
-5.2 V 50 mV
pp
-12 V 50 mV
pp
-24 V 150 mV
pp