1-6
Input C (Option 41)
Frequency Range:
40 MHz to 1.3 GHz
Sensitivity (Sinewave):
25 mV rms to 1 GHz
50 mV rms to 1.3 GHz
Dynamic Range:
40 dB to 1 GHz
Input Impedance:
50 ohms (nominal)
VSWR:
< 2:1 at 1GHz
Maximum Operating Input:
1 V rms
Maximum Input Without Damage:
7 V rms
1.5.2 Trigger Conditions
Automatic or manual selection in the +5.1 V
to -5.1 V range (X1), +51 V to -51 V range
(X10)
Trigger Level Range:
X1: ± 5.1 V in 20 mV steps
X10:± 51 V in 200 mV steps
Trigger Level Accuracy:
X1: ± 1% of Trigger Level + 30 mV
X10: ± 1% of Trigger Level + 300 mV
Autotrigger Minimum Amplitude:
150 mV pp
Autotrigger Frequency Range:
DC, 50 Hz to 160 MHz (typically to 200MHz)