1-8
1.5.3.4 Time Interval
Separate Inputs:
Input A start / Input B stop
Input B start / Input A stop
Common Inputs:
Input A start / Input A stop
Range:
-2 ns to 1 x 10 sec
5
Trigger Slopes:
Start, positive or negative
Stop, positive or negative
LSD:
1 ns (100 ps with average mode)
Resolution:
± LSD ± 1 ns ± Start Trigger Error* ± Stop
Trigger Error*
Accuracy:
± Resolution ± (Timebase Error x Time
Interval) ±Trigger Level Setting Error* ± 2 ns
1.5.3.5 Totalize A By B
Range:
100 MHz; 1 to 10 - 1 events
12
Pulse Width:
5 ns min at trigger points
Maximum Rate:
10 events / sec
8
Start / Stop:
Input B
LSD:
±1 count
Resolution:
LSD
Accuracy:
LSD
* Refer to Notes/Definitions