1-10
Accuracy:
±Resolution ± Trigger Level Timing Error* ±
Trigger Level Setting Error* at 10% Trigger
point ± Trigger Level Setting Error* at 90%
Trigger Point ± 2 ns ± (Timebase Error x
Rise/Fall Time)
1.5.3.9 Pulse Width
Range:
5 ns to 20 ms
Positive Pulse Width:
Start: + slope, 50% trigger point
Stop : - slope, 50% trigger point
Negative Pulse Width:
Start: - slope, 50% trigger point
Stop : + slope, 50% trigger point
Input Channel:
Input A
Minimum Pulse Height:
150 mV pp
LSD:
1 ns (100 ps using averaging mode)
Resolution:
± LSD ± 1 ns ± Start Trigger Error* ± Stop
Trigger Error*
Accuracy:
± Resolution ± Trigger Level Timing Error* ±
Trigger Level Setting Error* ± 2 ns ±
(Timebase Error x Pulse Width)
1.5.3.10
Phase A Rel B
Range:
0.1 to 360
0
LSD:
0.1 to 1 MHz
0
1 to 10 MHz
0
10 to 100 MHz
0
Resolution:
± LSD±(TI Resolution x 360
E
)/Period A
Accuracy
± LSD±(TI Resolution x 360
E
)/Period A
* Refer to Notes/Definitions
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com