4-5
4.1.7
Input B Sensitivity Test
4.1.7.0.1
Set the 2251A to its home state. Select 50 ohms input impedance for input B,
Frequency B as the measurement mode and external frequency standard.
(Command String: IP FB B1 BLI)
If Option 11 is installed connect the 2251A internal reference output to the external
signal source 10 MHz Input. Configure the external signal source to accept the
10MHz reference from the 2251A. (Command String: IP FB BLI)
4.1.7.1
Connect the Function Generator to the 2251A as shown in Figure 4-3.
Figure 4-3, Input B Low Frequency Sensitivity Test Configuration
4.1.7.2
Apply the frequencies given in Table 4-3 to the counters input B. Verify that the
2251A gives a stable reading within the tolerances given at less than the signal level given.
Frequency
Signal level
Resolution
Tolerance
10 Hz
25 mV rms 8 ± 1 Hz
5 kHz
25 mV rms 8
± 0.1 Hz
10 kHz
25 mV rms 8
± 0.1 Hz
Table 4-3, Input B Sensitivity Low Frequency Performance Limits
4.1.7.3
Disconnect the test equipment.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com