5-5
BNS
The 2251A will change to Positive Width (PW) mode and return a
header of PW. Channel A will have positive slope enabled and
Channel B will have negative slope enable.
Rise/Fall Time
The default settings, in addition to the above common settings, when the Rise Time (RT)
command is sent to the 2251A are:
COMMAND
Parameter Set
TI
Time Interval
APS
Positive Slope, Channel A
BPS
Positive Slope, Channel B
ADC
Channel A, DC Coupled
The following commands, when sent to the 2251A while Rise Time (RT) is enabled will
cause:
COMMAND
Parameter Set
ANS
The 2251A will change to Fall Time (FT) with a header of FT,
Channels A and B will have a negative slop enabled.
BNS
The 2251A will remain in the Rise Time (RT) mode with a header of
RT and Channels A and B with positive slope enabled.
The default settings, in addition to the above common settings, when the Fall Time (FT)
command is sent to the 2251A are:
COMMAND
Parameter Set
TI
Time Interval
ANS
Negative Slope, Channel A
BNS
Negative Slope, Channel B
ADC
Channel A, DC coupled
The following commands, when sent to the 2251A while Fall Time (FT) is enabled will cause:
COMMAND
Parameter Set
APS
The 2251A will change to Rise Time (RT) with a header of RT,
Channels A and B will have a positive slope enabled.