2461Type C User Manual
Racal Instruments
1998
Performance Verification and Adjustments 6-5
Table 6-3, Input 2 Sensitivity Low Frequency Tests
Frequency Signal
level Resolution
Command
Syntax
Tolerance
10Hz
25mV rms
SENS:RES 8
± 1Hz
5kHz
25mV rms
SENS:RES 8
± 0.1Hz
10kHz
25mV rms
SENS:RES 8
± 0.1Hz
Set the UCT measurement resolution to the values given in the
following table and apply the Frequencies given to the UCT
input 2. Check that the UCT gives a stable reading within the
tolerances given, at less than the signal levels given.
Table 6-4, Input 2 Sensitivity High Frequency Tests
Frequency Signal
level Resolution
Command
Syntax
Tolerance
100MHz
50mV rms
SENS:RES 9
± 1Hz
100kHz
50mV rms
SENS:RES 9
± 1Hz
10MHz
25mV rms
SENS:RES 8
± 1Hz
100kHz
25mV rms
SENS:RES 8
± 0.1Hz
Disconnect the test equipment.