2461Type C User Manual
Performance Verification and Adjustments 6-16
Racal Instruments
1998
INP1:FILT:LPAS:STAT ON
turn on LPF for channel 2
MEAS2:FREQ?
Confirm
50kHz
MEAS2:VOLT:MAX?
V
P2
MEAS2:VOLT:MIN?
V
N2
Use the equation :
or
to verify compliance.
(
) (
)
(
)
0
.
2
/
log
*
20
2
2
1
P1
>
−
−
N
P
N
V
V
V
V
(
) (
)
(
)
25
.
1
/
2
2
1
1
>
−
−
N
P
N
P
V
V
V
V
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com