User Manual 2461 Type C
A-6
×
1 Attenuation:
Selected if positive and negative peak amplitudes are less than
±
4.6V and the peak to peak amplitude of the signal does not
exceed 4.4V.
×
10 Attenuation:
Selected if positive or negative peak amplitudes are greater than
±
5.1V or the peak to peak amplitude of the signal exceeds 5.1V.
Auto-Trigger Minimum Amplitude
150mV pk-pk.
Auto-Trigger Level Accuracy
(Relative to true trigger point)
×
1
±
30mV
±
1% of trigger level reading.
×
10
±
300mV
±
1% of trigger level reading.
Frequency 1 or 2
Range
Channel 1 6 x 10
-4
Hz to 200MHz (Prescaled by 2)
Channel 2 6 x 10
-4
Hz to 100MHz
LSD Displayed
LSD = F x 10
-D
Where:
D = Number of digits selected.
F = frequency rounded to nearest decade.
Resolution
(selectable between 3 and 10 digits.)
For 3 to 5 Digits: