User Manual 2461 Type C
A-12
LSD
1ns (100ps with averaging)
Resolution
±
LSD
±
1ns RMS
±
START TRIGGER ERROR
ii
±
STOP TRIGGER ERROR
ii
Accuracy
±
RESOLUTION
±
(TIMEBASE ERROR x Wp)
±
(TRIGGER LEVEL TIMING ERROR
iii
)
±
2ns (DIFFERENTIAL CHANNEL DELAY ERROR
iv
)
where Wp = measured pulse width
Rise/Fall Time
Measurement Range
20ns to 20ms
Input Channel
INPUT 1 commoned with INPUT 2
Minimum Pulse Height
500mV pk-pk
Minimum Pulse Width
20ns at signal peaks
Rise Time Trigger Points
START on +ve slope at 10% of measured pk-pk