2461Type C User Manual
Programming The Type C Function Card 4-12
Racal Instruments
1998
Enables the math post-processing.
INTERROGATIVE FORM:
CALCulate:STATe?
==>
<OFF | ON>
Queries the state of the math post-processing.
MeASUREMENT
Subsystem
CONFigure
Subsystem
The commands in the CONFigure subsystem perform the
required configuration of the 2461 TYPE C instrument. The
SCALar commands program the 2461 TYPE C for one
measurement. The ARRay commands program the counter for
multiple measurements as specified by the <array size>
parameter. The limit for this parameter is 16384.
The input settings are not changed by these commands. These
must be set using the commands in the INPut subsystem.
The reconfiguration of the measurement function invalidates the
previously acquired data.
If the CONFigure command is sent while the 2461 TYPE C is
initiated, then the measurement in progress is aborted before
the reconfiguration takes place.
The CONFigure commands accept the expected value and
required resolution parameters. If both these values are
specified, and if they are valid, the number of digits of resolution
is determined.
CONFigure command summary
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com