2461Type C User Manual
Programming The Type C Function Card 4-14
Racal Instruments
1998
The ARRay commands program the counter for multiple
measurements as specified by the <array size> parameter.
The execution of these commands forces the following
settings: Auto Trigger is automatically enabled on the
selected channel and Common is deselected. The channel
trigger level is not updated so when these commands are
deselected, the trigger status returns to its previous state.
The optional <expected value> and <resolution> parameters
are used to derive the required number of digits of resolution.
Their units are volts.
INITIALIZATION:
*RST Resolution = 8
Array
size
=
1
EXAMPLE:
″
CONFigure1:VOLTage:MAXimum
″
Configures the 2461 TYPE C to measure the maximum
(positive peak) trigger level on channel 1 (single
measurement).
NO INTERROGATIVE FORM
*******************************************************
CONFigure[1 | 2 ][:SCALar]:FREQuency [<expected value>,
<resolution>]
CONFigure[1 | 2 ]:ARRay:FREQuency <array size> [,<expected
value>, Resolution>]
DESCRIPTION:
These commands configure the 2461 TYPE C to measure
the frequency of the signal on the input channel specified.
The ARRay command programs the counter for multiple
frequency measurements as specified by the <array size>
parameter.
The optional <expected value> and <resolution> parameters
are used to derive the required number of digits of resolution.
Their units are Hertz.
INITIALIZATION:
*RST Resolution = 8
Array
size
=
1