3152B User Manual
Publication No. 980935 Rev. A
5-14 Programming Reference
EADS North America Test and Services
Table 5-2, 3152B SCPI Command Summary
Keyword
Parameter Form
Default
3152A
HS
Instrument Control Commands
:INSTrument
:MODE
NORMal
|
FAST
NORM
9
0150A
[:SELect]
1 | 2
1
0008A
:COUPle
:MODE
MASTer
|
SLAVe
MAST
0002A
:DELay
0
to
20
0
0003A
:PATH
ADJacent | ECLT | LBUS
ADJ
0191A
:SLAVe
:DELete
<LAN_IP_address>
0040
:INSert
<3152B>,<LAN_IP_address>
00051
:STATe
OFF | ON | 0 | 1
0
0006A
Run Mode Commands
:INITiate
9
[:IMMediately]
9
:CONTinuous
OFF | ON | 0 | 1
1
9
1110A
:TRIGger
9
[:IMMediate]
9
11000
:BURSt
9
[:STATe]
OFF | ON | 0 | 1
0
9
1120A
:COUNt
1
to
1000000
1
9
1121A
:DELay
0, 10 to 2e6 (even numbers)
0 (0 = OFF)
9
1131A
:STATe
OFF | ON | 0 | 1
0
0188A
:TIMe
100e-9
to
20
100
0156A
:GATE
9
:MODe
LEVel
|
TRANsition
LEV
0157A
[:STATe]
OFF | ON | 0 | 1
0
9
1150A
:LEVel
-10 to 10
0/1.6 (*)
9
1180A
:SOURce
9
[:ADVance]
EXTernal | INTernal | TTLTrg<n> | ECLTrg1 | BUS
EXT
9
1140A
:SLOPe
POSitive | NEGative
POS
9
1160A
:RETRigger
[:STATe]
OFF | ON | 0 | 1
0
0187A
:TIMe
100e-9
to
20
100
0161A
:TIMer
1e-6 to 20
15e-6
9
1170A
(*) Legacy Mode defaults to 0 V; Modern Mode defaults to 1.6 V
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com