3152B User Manual
Publication No. 980935 Rev. A
7. Using ArbConnection prepare and download the following
waveform:
Segment 1:
Sine, 1000 points
Segment 2:
Triangle, 1000 points
Segment 3:
Square, 1000 points
Segment 4:
Sinc, 1000 points
Segment 5:
Gaussian Pulse, 1000 points
8. Using ArbConnection, build and download the following
sequence table:
Step 1:
Segment 1, loop 1
Step 2:
Segment 2, loop 1
Step 3:
Segment 3, loop 1
Step 4:
Segment 4, loop 1
Step 5:
Segment 5, loop 1
Test Procedure
1. From ArbConnection, click on the Manual Trigger and
observe that the waveforms advance through the sequence
table repeatedly.
Test Results
Pass
Fail
Note
Leave the same setup for the next test
Single Advance
Equipment: Oscilloscope, function generator
Preparation: (Same preparation as for previous step, except change
mode to single sequence advance)
1. Change Oscilloscope configuration to single
Test Procedure
1. From ArbConnection, click on the Manual Trigger and
observe that one waveform cycle advances through the
sequence table repeatedly with each external trigger signal.
Note that you might need to reset single-shot capture mode
on the oscilloscope for each trigger advance.
Test Results
Pass
Fail
6-18 Performance Checks