Publication No. 980935 Rev. A
3152B User Manual
FM - Standard
Waveforms
Equipment: Oscilloscope
Preparation:
1. Configure the oscilloscope as follows:
Time Base:
50
μ
s
Sampling Rate: 50 MS/s at least.
Trace A View: Jitter, Type: FREQ, CLK.
Trigger source: Channel 2, positive slope
Amplitude:
1 V/div
2. Connect the 3152B output to the oscilloscope input, channel 1.
3. Connect the 3152B SYNC to the oscilloscope input, channel 2.
4. Configure model 3152B controls as follows:
Waveform:
Modulated
Modulation:
FM
Carrier Freq:
1 MHz
Mod Frequency: 10 kHz
Deviation: 500
kHz
Sync: On
Output: On
Test Procedure:
1. Verify FM operation on the oscilloscope as follows:
Waveform: Sine
Frequency: 10
kHz
Max F:
1.25 MHz
Min F:
750 kHz
Test Results
Pass
Fail
2. Modify the 3152B modulating waveform to triangle, then square,
and then ramp to verify all standard FM waveforms.
Test Results
Pass
Fail
3. Move the 3152B marker position to 1.25MHz and verify the
marker position.
Test Results
Pass
Fail
Triggered FM -
Equipment: Oscilloscope, function generator
EADS North America Test and Services
Performance Checks 6-25