3152B User Manual
Publication No. 980935 Rev. A
(Setup 50MHz)
50 MHz Gated Oscillator Adjustment
Equipment:
Counter, Function Generator, BNC to BNC cables
Preparation:
1. Configure the counter as follows:
Termination:
50
Ω
DC
Function:
TI A -> B
Slope
B:
Negative
2. Connect the 3152B output to the oscilloscope input
3. Connect an external function generator to the front panel
TRG/PLL connector
4. Using ArbConnection, prepare and download the following
waveform:
Wavelength:
100
points
Waveform:
Pulse: Delay = 0.01 %,
Rise/Fall Time = 0 %, High Time = 99.99 %
5. Configure the 3152B as follows:
Function
Mode: Arbitrary
Run
Mode:
Triggered
Retrigger
Mode: On
Retrigger Delay: 20
μ
s
6. Using an external function generator, manually trigger the
3152B
Adjustment:
1. Set CAL:SETUP 0
2. Adjust C10 for a period of 20
μ
s
±
5%
Setup TCXO
10 MHz TCXO Frequency
Equipment:
Counter, BNC to BNC cables
Preparation:
1. Configure the counter as follows:
Function:
Freq
A
Termination:
50
Ω
2. Connect the 3152B output to counter channel A.
3. Configure the 3152B as follows:
Frequency:
10
MHz
Ref
Source:
Internal
Output:
On
Amplitude
2
V
Wave:
Square
Adjustment:
1. Adjust CAL:SETUP 1 for a counter reading of 10 MHz
±
2 Hz
7-10 Adjustments and Firmware Update