3152B User Manual
Publication No. 980935 Rev. A
Setup 54
100 mV Amplitude, Post Attenuators - Modulation
Equipment: DMM, BNC to BNC cable, 50
Ω
feedthrough termination,
Dual banana to BNC adapter
Preparation:
1. Configure the DMM as follows:
Function:
ACV
Range:
100
mV
2. Connect the 3152B output to the DMM input. Terminate the
3152B output at the DMM input with the 50
Ω
feedthrough
termination.
3. Configure the 3152B as follows:
Frequency:
1
kHz
Output:
On
Amplitude:
100
mV
Mode:
Modulation
Adjustment:
1. Adjust CAL:SETUP 38 for a DMM reading of 35.35 mV
±
0.5 mV
Setup 55
50 mV Amplitude, Post Attenuators - Modulation
Equipment: DMM, BNC to BNC cable, 50
Ω
feedthrough termination,
Dual banana to BNC adapter
Preparation:
1. Configure the DMM as follows:
Function:
ACV
Range:
100
mV
2. Connect the 3152B output to the DMM input. Terminate the
3152B output at the DMM input with the 50
Ω
feedthrough
termination.
3. Configure the 3152B as follows:
Frequency:
1
kHz
Output:
On
Amplitude:
50
mV
Mode:
Modulation
Adjustment:
1. Adjust CAL:SETUP 39 for a DMM reading of 17.67 mV
±
0.5 mV
7-28 Adjustments and Firmware Update
EADS North America Test and Services
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com