User Manual 3162A/B
Using The Instrument 3-15
The default output function of the generator is the sine
waveform. Since the instrument uses the filter to generate
this waveform, the state of the OUTP:FILT cannot be
changed until another output function is selected. A
setting conflict error will occur if one attempts to change
the filter state before changing to another output function
or mode.
The WaveCAD Filter Control Panel is shown below as a graphical
implementation of the filter setting.
Figure 3-4, WaveCAD FilterControl Panel
The query:
OUTPut:FILTer?
queries the output filter state and returns "0" (OFF) or "1" (ON).
Activating the
Backplane ECLTRG
and TTLTRG Lines
The Model 3162B can transmit and receive signals on VXIbus ECLTRG line
zero and TTLTRG lines zero through three.
Use the following command to activate one of two backplane ECLTRG lines:
OUTPut:ECLTrg0 { OFF | ON}
ON enables the ECL trigger line. The trigger source for this line can be
selected with the TRIG:SOUR command. The default ECLTrg0 state is OFF.
Note that the trigger validation source for ECLT0 is exactly as was set for the
SYNC output.
Summary of Contents for 3162A/B
Page 4: ...This page was left intentionally blank...
Page 10: ...3162A B User Manual vi Support Offices 6 2...
Page 12: ...3162A B User Manual viii This page was left intentionally blank...
Page 17: ...User Manual 3162A B Getting Started 1 5 Figure 1 2 The Front Panel...
Page 23: ...User Manual 3162A B Getting Started 1 11 Figure 1 7 WaveCAD Example...
Page 27: ...User Manual 3162A B Getting Started 1 15 This page was left intentionally blank...
Page 28: ......
Page 34: ...3162A B User Manual Configuring The Instrument 2 6 This page was left intentionally blank...
Page 78: ......