OPERATION AND TREATMENT PROTOCOLS
57
Mistral User Manual
•
Have a history of herpes outbreaks in the treatment area,
unless the patient has consulted their physician and received
prophylaxis treatment prior to treatment.
•
Have a history of skin cancer or patients who have areas of
potential skin malignancies.
•
Have received chemotherapy or radiotherapy treatments
within the past 5 years should not be treated without their
acting physician's consent.
•
Have been on a steroid regimen (i.e. asthma) for the last 3
months.
•
Have an active implant, such as a pacemaker, incontinence
device, insulin pump, etc.
•
Have any other condition which in the practitioner’s opinion
would make it unsafe for the patient to be treated.
4.7.4.
Pre-Treatment Test – Acne Clearance
Set the energy level for the first pulse according to the table below. For
example, if the patient is skin Type II, the energy level for the first test
pulse should be set at 20.
Table 10: Energy Level Setting - AC
I
II
III
IV
V
VI
1st Pulse
25
20
15
10
3
0
2nd Pulse
30
25
20
15
5
3
3rd Pulse
35
30
25
20
8
5
The appropriate energy setting is the highest possible energy level that
causes mild redness to the skin and disappears within a few hours
after test.
If the required reaction in the target area is observed, this is the
appropriate energy level for treatment. If this required result does not
appear increase the energy level according to Table 10 until the