For transmitters rated at a maximum output power not listed above, the recommended separation distance 

d

 in meters (m) can 

be estimated using the equation applicable to the frequency of the transmitter, where p is the maximum output power rating of 
the transmitter in watts (W) according to the transmitter manufacturer. 
NOTE1    At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. 
NOTE2    These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection 
from structures, objects and people.

 

Manufacturer’s declaration-electromagnetic immunity 

Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment 

The  TH8xyz series  is intended for use in the electromagnetic environment (for home healthcare) specified below. 
The customer or the user of the  TH8xyz series  should assure that it is used in such an environment.

 

Test 

frequency 

(MHz) 

Band 

a)

 

(MHz) 

Service 

a)

 Modulation 

b)

 

Maximum 

power 

(W) 

Distance 

(m) 

IMMUNITY TEST 

LEVEL 

(V/m) 

Compliance 

LEVEL 

(V/m) 

(for home 

healthcare) 

385 

380 –390 

TETRA 400 

Pulse 

modulation b) 

18 Hz 

1,8 0,3 

27 

27 

450 

430 – 470 

GMRS 460, 

FRS 460 

FM c) 

±5 kHz deviation

 

1 kHz sine 

2 0,3  28 

28 

710 

704 – 787 

LTE Band 13, 

17 

Pulse 

modulation b) 

217 Hz 

0,2 0,3 

745 

780 

810 

800 – 960 

GSM 800/900, 

TETRA 800, 

iDEN 820, 

CDMA 850, 

LTE Band 5 

Pulse 

modulation b) 

18 Hz 

2 0,3  28 

28 

870 

930 

1 720 

1 700 – 

1 990 

GSM 1800; 

CDMA 1900; 

GSM 1900;   

DECT; 

LTE Band 1, 3, 

4, 25; UMTS 

Pulse 

modulation b) 

217 Hz 

2 0,3  28 

28 

1 845 

1 970 

2 450 

2 400 – 

2 570 

Bluetooth,  

WLAN, 

802.11 b/g/n,   

RFID 2450,   

LTE Band 7 

Pulse 

modulation b) 

217 Hz 

2 0,3  28 

28 

5 240 

5 100 – 

5 800 

WLAN 802.11 

a/n 

Pulse 

modulation b) 

217 Hz 

0,2 0,3 

5 500 

5 785 

NOTE    If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and 
the ME EQUIPMENT or ME  SYSTEM may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3. 

a)          For some services, only the uplink frequencies are included. 
b)          The carrier shall be modulated using a 50 % duty cycle square wave signal. 

c)          As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because while it does 

not  represent actual modulation, it would be worst case. 

 

 