Dynamic dose
During Incremental IP titration the size of the increments is controlled
by the Dynamic dose. Small increments will be added in the vicinity of
the curve inflections, and larger increments will be added at the flat
part of the curve. The size of the increments cannot be greater than
the Maximum dose.
The higher the value is, the larger the increments will be. Decreasing
the Dynamic dose value slows down the titration, but leads to more
accurate titration. However, too small a value can lead to the detection
of false inflection points.
The standard value, 10, will be appropriate for standard
applications. If you want to use increments of the same size,
select an IP Monotonic method.
Enter in:
Method parameters menu of a Dynamic IP method.
Range available:
1 to 999