Titra
Lab
®
TitraLab 870 Reference Manual
Page 102
Dynamic rinses
Dynamic rinses are performed by a SAC850 or SAC950 Sample
Changers if the Dynamic rinsing module is installed on the sample
changer.
A Dynamic rinse performed in a Park or Rinse beaker consists of the
following operations:
1.
The electrode are positioned above the Park or Rinse beaker.
2.
The electrodes are dipped into the beaker. The beaker is emptied
to a waste in the same time. At the end, the electrodes are
located in the emptied beaker at their downmost position.
3.
The electrodes are washed with rinse solution (usually demineral-
ised water) then start to move up under rinsing. At the end, the
electrodes are located above a beaker filled with rinse solution
and some remaining impurities.
4.
Steps
2
and
3
can be repeated up to 8 times as up to 9 dynamic
rinses can be programmed. Steps
2
and
3
are performed under
stirring.
A Dynamic rinse performed in a Sample beaker consists of the follow-
ing operations:
1.
The electrodes are dipped into the Sample beaker. The beaker is
emptied to a waste in the same time. At the end, the electrodes
are located in the emptied beaker at their downmost position.
2.
The electrodes move up and are washed in the same time with
rinse solution (usually demineralised water). At the end, the elec-
trodes are located above a beaker filled with rinse solution and
some remaining impurities.
3.
The sample beaker is emptied to a waste.
4.
A last rinsing of the electrodes and delivery tips is carried out. At
the end, the electrodes are located above a nearly emptied
beaker containing a little solvent that was used to flush the elec-
trodes and delivery tips.
5.
Steps
1
to
3
can be repeated up to 8 times as up to 9 dynamic
rinses can be programmed. Steps
1
to
3
are performed under
stirring.
E0 standard
potential
Refer to "Direct ISE measurement method - definition", page 95
Summary of Contents for TitraLab 870
Page 1: ...TitraLab 870 pH EP IP EC ISE Titration Workstation Reference Manual D21T060...
Page 2: ...D21T060 Printed by Radiometer Analytical SAS 2008 02F...
Page 22: ...TitraLab TitraLab 870 Reference Manual Page 22...
Page 23: ...Practical examples s...
Page 24: ......
Page 32: ...TitraLab TitraLab 870 Reference Manual Page 32...
Page 40: ...TitraLab TitraLab 870 Reference Manual Page 40...
Page 41: ...Glossary...
Page 42: ......
Page 255: ...Appendixes...
Page 256: ......
Page 258: ...TitraLab TitraLab 870 Reference Manual Page 258...
Page 262: ...TitraLab TitraLab 870 Reference Manual Page 262...
Page 280: ...TitraLab TitraLab 870 Reference Manual Page 280...
Page 286: ...TitraLab TitraLab 870 Reference Manual Page 286...