Electrode
window
This window contains all the information and operations concerning
the electrodes.
To access:
Use
LEFT/RIGHT
arrow keys.
Electrodes are
different
This message appears if the active electrode used for the reagent
calibration is different from the electrode used in the method.
Check the method and reagent calibration parameters menus.
Empty burette
1.
Enter the Burette functions menu.
2.
Select the burette in the
Address
.
3.
Press
2
Empty
. The burette is emptied at maximum speed
(approximate emptying time: 25 s).
Empty
sequence
Involves removing all the methods present in the sequence.
Proceed as follows:
1.
Select the sequence to be emptied,
.
2.
Press
2
Sequence/Sample stack then
3
Edit sequence.
3.
Press
3
Delete.
4.
Press
2
Delete sequence then
✓
to confirm or press
Esc
to leave
the screen without emptying the sequence.
To remove a particular method from a sequence,
Summary of Contents for TitraLab 870
Page 1: ...TitraLab 870 pH EP IP EC ISE Titration Workstation Reference Manual D21T060...
Page 2: ...D21T060 Printed by Radiometer Analytical SAS 2008 02F...
Page 22: ...TitraLab TitraLab 870 Reference Manual Page 22...
Page 23: ...Practical examples s...
Page 24: ......
Page 32: ...TitraLab TitraLab 870 Reference Manual Page 32...
Page 40: ...TitraLab TitraLab 870 Reference Manual Page 40...
Page 41: ...Glossary...
Page 42: ......
Page 255: ...Appendixes...
Page 256: ......
Page 258: ...TitraLab TitraLab 870 Reference Manual Page 258...
Page 262: ...TitraLab TitraLab 870 Reference Manual Page 262...
Page 280: ...TitraLab TitraLab 870 Reference Manual Page 280...
Page 286: ...TitraLab TitraLab 870 Reference Manual Page 286...