Equivalent point
determination
(IP methods)
Incremental addition of the titrant (Monotonic/Dynamic IP
methods):
•
acquisition of the titration curve E (electrode) = f (volume of titrant
dispensed),
•
determination of the first derivative curve using a smoothing proc-
ess (depending on the value set for IP filter),
•
determination of the second derivative curve,
•
determination of the maxima of the first derivative curve using the
IP filter parameter. These maxima correspond to the zeros of the
second derivative curve.
•
determination of the equivalent point: for this point the maximum of
the first derivative curve must be equal to or greater than the preset
IP reject method parameter.
Figure 23: Determination of an equivalent point - Monotonic or Dynamic IP
method
An equivalent point is determined and plotted live on the titration
curve. The method is stopped after a given number of equivalent
points has been detected.
E or pH
Summary of Contents for TitraLab 870
Page 1: ...TitraLab 870 pH EP IP EC ISE Titration Workstation Reference Manual D21T060...
Page 2: ...D21T060 Printed by Radiometer Analytical SAS 2008 02F...
Page 22: ...TitraLab TitraLab 870 Reference Manual Page 22...
Page 23: ...Practical examples s...
Page 24: ......
Page 32: ...TitraLab TitraLab 870 Reference Manual Page 32...
Page 40: ...TitraLab TitraLab 870 Reference Manual Page 40...
Page 41: ...Glossary...
Page 42: ......
Page 255: ...Appendixes...
Page 256: ......
Page 258: ...TitraLab TitraLab 870 Reference Manual Page 258...
Page 262: ...TitraLab TitraLab 870 Reference Manual Page 262...
Page 280: ...TitraLab TitraLab 870 Reference Manual Page 280...
Page 286: ...TitraLab TitraLab 870 Reference Manual Page 286...