TitraLab 870 Reference Manual
Page 157
Min. speed -
Max. speed
This is the minimum/maximum speed at which the burette will deliver
titrant during an analysis with continuous addition of the titrant (end
point titration for example).
The
Min. speed
must always be a low value in order to obtain
accurate results. A value of 0.2 ml/min. can be entered for most appli-
cations. A low value entered for the
Max. speed
will slow down
the titration whereas too high a value will result in an overshoot of the
equivalent point.
Enter in:
Method parameters menu
Reagent parameters menu
Range available:
Min. speed = 0.01 ml/min. to Max. speed
Max. speed = Min. speed to 50.00 ml/min.
Refer to "Burette speed", page 69
Min. Temp. -
Max. Temp.
Acceptance range for the temperature measured in the standard
solution. If the temperature lies outside the defined range the calibra-
tion is stopped.
Enter in:
Edit electrode > Result menu
Range available:
Min. Temp. = -9°C to + Max. Temp.
Max. Temp = Min. Temp. to +99°C
Summary of Contents for TitraLab 870
Page 1: ...TitraLab 870 pH EP IP EC ISE Titration Workstation Reference Manual D21T060...
Page 2: ...D21T060 Printed by Radiometer Analytical SAS 2008 02F...
Page 22: ...TitraLab TitraLab 870 Reference Manual Page 22...
Page 23: ...Practical examples s...
Page 24: ......
Page 32: ...TitraLab TitraLab 870 Reference Manual Page 32...
Page 40: ...TitraLab TitraLab 870 Reference Manual Page 40...
Page 41: ...Glossary...
Page 42: ......
Page 255: ...Appendixes...
Page 256: ......
Page 258: ...TitraLab TitraLab 870 Reference Manual Page 258...
Page 262: ...TitraLab TitraLab 870 Reference Manual Page 262...
Page 280: ...TitraLab TitraLab 870 Reference Manual Page 280...
Page 286: ...TitraLab TitraLab 870 Reference Manual Page 286...