Number of
additions
Set the number of additions to be performed automatically by an
Addition method or an ISE standard addition method.
Refer to "Addition method - definition", page 48
Refer to "ISE Standard addition method - definition", page 142
Enter in:
Edit method menu (Addition method)
Edit method > Parameters (ISE standard addition method)
Range available:
1 to 3 (Addition method)
1 to 9 (ISE standard addition method)
Number of
buffers
Available if Calibration request = Yes
Number of pH standards to be used for the calibration. Work with at
least two standards to calculate the electrode sensitivity. If one
standard is only used the zero pH is calculated.
Enter in:
Edit electrode menu
Range available:
1 to 5
Refer to "pH buffer", page 169
Number of
cycles
Available if Calibration request = Yes
Number of times the calibration is to be repeated, i.e. the number of
beakers to be prepared for each pH, conductivity or ISE standard.
Enter in:
Edit electrode menu
Range available:
1 to 9
Number of
decimals
Number of decimals (0 to 3) to be displayed and printed for the result.
Enter in:
Edit method > Results (for a pH measurement method)
Summary of Contents for TitraLab 870
Page 1: ...TitraLab 870 pH EP IP EC ISE Titration Workstation Reference Manual D21T060...
Page 2: ...D21T060 Printed by Radiometer Analytical SAS 2008 02F...
Page 22: ...TitraLab TitraLab 870 Reference Manual Page 22...
Page 23: ...Practical examples s...
Page 24: ......
Page 32: ...TitraLab TitraLab 870 Reference Manual Page 32...
Page 40: ...TitraLab TitraLab 870 Reference Manual Page 40...
Page 41: ...Glossary...
Page 42: ......
Page 255: ...Appendixes...
Page 256: ......
Page 258: ...TitraLab TitraLab 870 Reference Manual Page 258...
Page 262: ...TitraLab TitraLab 870 Reference Manual Page 262...
Page 280: ...TitraLab TitraLab 870 Reference Manual Page 280...
Page 286: ...TitraLab TitraLab 870 Reference Manual Page 286...