TitraLab 870 Reference Manual
Page 209
Results by
difference/
cumulate
Results by difference (ml)
Volumes between two End points or Inflection points are used for
result calculation.
Results by cumulate (ml)
Total volume delivered up to the End point or Inflection point is used
for result calculation.
This setting has no influence on the result calculation for the
first equivalence point detected.
Enter in:
Edit method > Results (End point, IP monotonic, IP dynamic or IP
continuous methods)
Results factor
(Yes/No)
This multiplying factor is applied to all Rn (n = 1 to 8) results of the
method. Equation results En (n = 1 or 2) are not processed.
Access:
Edit method > Results
Proceed as follows to multiply all your method results by a factor:
•
In the method Results menu, select
Results factor =
Yes
.
•
Run the method:
There is no results factor for a reagent calibration method.
Enter the results factor (from 0.001 to 10 000)
then press
1
to start the method.
Summary of Contents for TitraLab 870
Page 1: ...TitraLab 870 pH EP IP EC ISE Titration Workstation Reference Manual D21T060...
Page 2: ...D21T060 Printed by Radiometer Analytical SAS 2008 02F...
Page 22: ...TitraLab TitraLab 870 Reference Manual Page 22...
Page 23: ...Practical examples s...
Page 24: ......
Page 32: ...TitraLab TitraLab 870 Reference Manual Page 32...
Page 40: ...TitraLab TitraLab 870 Reference Manual Page 40...
Page 41: ...Glossary...
Page 42: ......
Page 255: ...Appendixes...
Page 256: ......
Page 258: ...TitraLab TitraLab 870 Reference Manual Page 258...
Page 262: ...TitraLab TitraLab 870 Reference Manual Page 262...
Page 280: ...TitraLab TitraLab 870 Reference Manual Page 280...
Page 286: ...TitraLab TitraLab 870 Reference Manual Page 286...