Titra
Lab
®
TitraLab 870 Reference Manual
Page 240
Std conc. too
low (or high)
This message is displayed when running an ISE standard addition
method with automatic determination of the volumes in the following 2
cases:
•
The TIM870 performs a first addition equal to 0.1% of the aliquot
volume. If after this first addition, the potential jump is already
higher than the maximum value (
∆
E total *(1+0,1*n) / n" mV
), the
Std conc. too high
error message is displayed.
∆
E total
and
n
are user-programmable parameters of an ISE
standard addition method :
Potential jump
(from 5 to 100 mV) and
Number of additions
(from 1 to 9).
•
If the cumulate addition volumes become higher than 1 % of the
aliquot volume while the potential jump measured remains below
1 mV
, the
Std conc. too low
error message is displayed.
In the 2 cases, the test is rejected and you can start a new test
(press
2
then
1
New test) or end the analysis (press
3
).
Summary of Contents for TitraLab 870
Page 1: ...TitraLab 870 pH EP IP EC ISE Titration Workstation Reference Manual D21T060...
Page 2: ...D21T060 Printed by Radiometer Analytical SAS 2008 02F...
Page 22: ...TitraLab TitraLab 870 Reference Manual Page 22...
Page 23: ...Practical examples s...
Page 24: ......
Page 32: ...TitraLab TitraLab 870 Reference Manual Page 32...
Page 40: ...TitraLab TitraLab 870 Reference Manual Page 40...
Page 41: ...Glossary...
Page 42: ......
Page 255: ...Appendixes...
Page 256: ......
Page 258: ...TitraLab TitraLab 870 Reference Manual Page 258...
Page 262: ...TitraLab TitraLab 870 Reference Manual Page 262...
Page 280: ...TitraLab TitraLab 870 Reference Manual Page 280...
Page 286: ...TitraLab TitraLab 870 Reference Manual Page 286...