TitraLab 870 Reference Manual
Page 50
Addition volume
Refer to "Reagent addition volume", page 189
.
Address
The position where the electrode and burette are placed during opera-
tion:
Electrode
The electrode address is defined using the format “xxx/y” where “xxx”
corresponds to the instrument (TIM, ABU1 or ABU2) where the
electrode is connected and “y” corresponds to the socket.
For example TIM/E1, indicates that the electrode is connected to E1
socket on the Titration Manager.
It is always recommended to connect an indicating electrode
and its associated reference to the same instrument (TIM,
ABU1 or ABU2).
Refer to "Electrode connection", page 114
Burette
The burette address is defined using the format “xxx/y” where “xxx”
corresponds to the instrument (TIM, ABU1 or ABU2) where the burette
is placed and “y” corresponds to the position. For example TIM/1, indi-
cates that the burette is placed on the Titration Manager in position 1.
Figure 2: Burette positions
.
Alarm: Locked
The user cannot bypass an electrode, reagent calibration and/or QC
sample analysis if the last result obtained lies outside the acceptance
range.
Enter in:
Setup menu > Access routine mode
Position 2
Position 1
Summary of Contents for TitraLab 870
Page 1: ...TitraLab 870 pH EP IP EC ISE Titration Workstation Reference Manual D21T060...
Page 2: ...D21T060 Printed by Radiometer Analytical SAS 2008 02F...
Page 22: ...TitraLab TitraLab 870 Reference Manual Page 22...
Page 23: ...Practical examples s...
Page 24: ......
Page 32: ...TitraLab TitraLab 870 Reference Manual Page 32...
Page 40: ...TitraLab TitraLab 870 Reference Manual Page 40...
Page 41: ...Glossary...
Page 42: ......
Page 255: ...Appendixes...
Page 256: ......
Page 258: ...TitraLab TitraLab 870 Reference Manual Page 258...
Page 262: ...TitraLab TitraLab 870 Reference Manual Page 262...
Page 280: ...TitraLab TitraLab 870 Reference Manual Page 280...
Page 286: ...TitraLab TitraLab 870 Reference Manual Page 286...