Titra
Lab
®
TitraLab 870 Reference Manual
Page 65
Beaker
detection
minimum height
The table below reports the minimum height of liquid that must be
present in the beaker in order to detect the beaker as not empty,
also "Beaker detection", page 64
Beaker type
diameter x height (mm)
Detection minimum
height
Part no.
400 ml tall form 70 x 131
10 mm (40 ml)
250 ml low form 70 x 95
10 mm (40 ml)
250 ml tall form 60 x 120
10 mm (30 ml)
125 ml, Gosselin, PP,
54 x 73
10 mm (25 ml)
X31T087 (pack of 50)
180 ml, Gosselin, PP,
54 x 103
10 mm (25 ml)
X31V005 (pack of 50)
150 ml tall form 60 x 80
10 mm (30 ml)
150 ml low form 62 x 81
10 mm (30 ml)
40-100 ml, PP 60 x 80
10 mm (30 ml)
904-490 (pack of 50)
100 ml tall form 48 x 80
10 mm (20 ml)
50 ml low form 42 x 60
10 mm (15 ml)
904-489 (pack of 50)
22-45 ml, PP 44 x 70
10 mm (15 ml)
904-489 (pack of 50)
8-25 ml, PP 44 x 70
10 mm (15 ml)
904-488 (pack of 50)
Summary of Contents for TitraLab 870
Page 1: ...TitraLab 870 pH EP IP EC ISE Titration Workstation Reference Manual D21T060...
Page 2: ...D21T060 Printed by Radiometer Analytical SAS 2008 02F...
Page 22: ...TitraLab TitraLab 870 Reference Manual Page 22...
Page 23: ...Practical examples s...
Page 24: ......
Page 32: ...TitraLab TitraLab 870 Reference Manual Page 32...
Page 40: ...TitraLab TitraLab 870 Reference Manual Page 40...
Page 41: ...Glossary...
Page 42: ......
Page 255: ...Appendixes...
Page 256: ......
Page 258: ...TitraLab TitraLab 870 Reference Manual Page 258...
Page 262: ...TitraLab TitraLab 870 Reference Manual Page 262...
Page 280: ...TitraLab TitraLab 870 Reference Manual Page 280...
Page 286: ...TitraLab TitraLab 870 Reference Manual Page 286...