TitraLab 870 Reference Manual
Page 86
Coupled
method
A Coupled method is a combination of methods performed in the
same beaker. When using a coupled method, the instrument runs all
these methods on the same sample.
If you wish to run a series of methods in different beakers, it is
necessary to program a Sequence instead of a Coupled method.
Example: Combination of method 1 and method 2.
The number of test portions (for example 3) is entered during pro-
gramming. The method is then repeated in the number of beakers
specified.
Figure 17: Coupled method with three tests
= beaker number 1
Sample 1
1
1
1
2
2
3
3
Test portion 1
Test portion 2
Test portion 3
Method 1
Method 2
Method 1
Method 2
Method 1
Method 2