Titra
Lab
®
TitraLab 870 Reference Manual
Page 172
Potential versus
SHE
This parameter is available when creating a reference or a combined
electrode with the option
From = Other
.
Refer to "Create electrode", page 87
This parameter enables you to calibrate electrodes with automatic
buffer checking using any kind of reference electrodes (for example
with a mercurous sulphate electrode Hg/Hg2SO4 (Sat. K2SO4)). You
just need to know the potential of the reference electrode versus the
Standard Hydrogen Electrode.
This potential is taken into account in the buffer recognition algorithm
and for pH calculation when no calibration is performed. When you
create a reference or a combined electrode with the option
From =
Catalogue
, the instrument calculates pH using the potential
versus SHE stored in memory for the reference electrode selected. In
this case, potential versus SHE cannot be changed.
The table below gives the potential at 25°C versus the SHE (E SHE)
of a few "reference elements/filling solution" couples.
Table 8: Reference electrode potentials versus SHE
Range available:
0.0 mV to +1000.0 mV
Reference element and filling
solution
E SHE
(mV)
Radiometer Analytical
Reference electrodes
Hg/Hg
2
Cl
2
– Sat. KCl
SCE: Saturated Calomel Electrode
+244
pHC4000, pHC4001, pHC4006,
XC601, REF401, REF421, REF451,
XR100, XR110, XR130, XR150,
MC408PtREF401, REF421, PHC4000
Hg/Hg
2
Cl
2
– 1M LiCl
+280
REF921
Hg/Hg
2
SO
4
- Saturated K
2
SO
4
+651
REF601,
REF621,
XR200,
XR230,
MC602Pt, MC6091Ag
Hg/Hg
2
SO
4
- 1 M H
2
SO
4
+616
Hg/HgO - 0.1 M KOH
+174
XR400, XR430, XR440
Ag/AgCl - Saturated KCl
+199
XR300, XR820, XC100, XC111,
XC120, XC161, XC200, XC250
Ag/AgCl - 3 M KCl
+208
pHC3001, pHC3005, pHC3006,
pHC3011, pHC3081, pHC3185,
REF321, REF361, MC3051Pt,
ISEC301F
Ag/AgCl - 1 M KCl
+235
Ag/AgCl - 0.6 M KCl (sea water)
+250
Red Rod - Saturated KCl
+199
pHC2001, pHC2002, pHC2003,
pHC2005, pHC2011, pHC2015,
pHC2051, pHC2085, pHC2401,
pHC2441, pHC2501, pHC2601,
pHC2701, REF200, REF201, REF251,
REF261, MC2095Sb,MC201Au-8
Summary of Contents for TitraLab 870
Page 1: ...TitraLab 870 pH EP IP EC ISE Titration Workstation Reference Manual D21T060...
Page 2: ...D21T060 Printed by Radiometer Analytical SAS 2008 02F...
Page 22: ...TitraLab TitraLab 870 Reference Manual Page 22...
Page 23: ...Practical examples s...
Page 24: ......
Page 32: ...TitraLab TitraLab 870 Reference Manual Page 32...
Page 40: ...TitraLab TitraLab 870 Reference Manual Page 40...
Page 41: ...Glossary...
Page 42: ......
Page 255: ...Appendixes...
Page 256: ......
Page 258: ...TitraLab TitraLab 870 Reference Manual Page 258...
Page 262: ...TitraLab TitraLab 870 Reference Manual Page 262...
Page 280: ...TitraLab TitraLab 870 Reference Manual Page 280...
Page 286: ...TitraLab TitraLab 870 Reference Manual Page 286...